Sapphire Infusion Pump User Manual189
> To repeat the last infusion:
1. From the Start Up screen, select Repeat Last Infusion.
2. Review the parameters displayed on the Confirm screen.
Then, press OK.
3. To begin the infusion, make sure that the clamps on the administration set are
open, and press Start.
The Running screen appears, and the infusion begins.
Using a PreSet Program
The PreSet Programs function allows you to select an infusion with predefined
parameters. Each delivery mode can support up to 10 PreSet programs. PreSet
programs are available only in the delivery mode in which they were saved.
Information about Secondary (Piggyback) infusions is not saved.
This option is not relevant to Secondary infusions.
Repeat Last Infusion is not available (grayed out) when the pump
settings differ from those used for the previous infusion. These
include: CCA, regional settings, and PCA/PCEA infusion type.
Repeat Last Infusion does not include a loading dose, even if one
was programmed for the original infusion. If required, a clinician
bolus can be given when the infusion starts. For more information
refer to Administering a Clinician Bolus on page 160.
Verify that the parameters reflect the correct treatment according to
the prescription.
For the PreSet Programs option to appear on the Start Up screen,
the PreSet programs need to be enabled on the pump. For more
information, refer to Configuring General Settings on page 204.