Technical Specifications
270
The following sections present technical specifications for the pump and its
components:
Pump Accuracy .................................................................................................... 270
Pump Specifications ............................................................................................. 276
Average Bolus Volume After Occlusion .................................................................. 277
Environmental Specifications ................................................................................ 278
Electromagnetic Compatibility Statement .............................................................. 281
Pump Accuracy
The following graphs and curves were derived from the pump accuracy testing
procedures described in the IEC60601-2-24 standard. Testing was performed under
normal conditions (specified in IEC60601-2-24 standard) at room temperature (25°C,
72°F).
Normal conditions to ensure optimal accuracy of ±2.5%:
Fluid level should be 50 cm above the pump
No back pressure due to catheter size or difference in height of pump and infusion
site
Room temperature (25°C)
Barometric pressure of sea level altitude (101kPa)
IV medication with water like fluid characteristics