MEDICA Pro 30/60/120 - US
Operator Manual

Questions and Answers

  • C
    Colleen LeeAug 16, 2025
    What to do if I detect a leak in my ELGA Laboratory Equipment?
    • D
      daniel15Aug 16, 2025
      If you detect a leak, immediately press the mute button to silence the alarm. Then, remove the power, rectify the leak, and dry the sensors. Ensure you also dry the contacts, and finally, power the system back on to reset.
  • M
    Mary FowlerAug 19, 2025
    What to do if the break tank is low in my ELGA MEDICA Pro 60 Laboratory Equipment?
    • C
      Christine HutchinsonAug 19, 2025
      If the break tank is low, start by pressing the mute button to silence the alarm. Next, confirm the pressure and flow of the potable supply. If the problem continues, you may need to contact a local service provider.
  • B
    butlerjeffreyAug 22, 2025
    How to resolve a water temperature alarm on my ELGA Laboratory Equipment?
    • C
      ccainAug 22, 2025
      If you receive a water temperature alarm, begin by pressing the mute button to silence the alarm. Next, confirm that the water temperature is suitable for your application. You may need to adjust the alarm setpoint or direct water to drain to introduce cool water.
  • A
    Amy CaseAug 26, 2025
    What should I do about the permeate temperature alarm on my ELGA MEDICA Pro 60?
    • T
      tclarkAug 26, 2025
      If the permeate temperature alarm is triggered, first press the mute button to silence the alarm. Then, confirm that the temperature is suitable for your application. You might need to adjust the alarm setpoint or confirm the potable supply temperature is suitable.
  • W
    wscottAug 28, 2025
    What to do if I have problems with reservoir level controls in my ELGA MEDICA Pro 60?
    • M
      Marcus KimAug 28, 2025
      If you are experiencing issues with the reservoir level controls, start by pressing the mute button to silence the alarm. Then, remove the power from the unit. If the problem persists after these steps, it is recommended to contact Technical Support.
  • M
    Michael NelsonAug 31, 2025
    What to do if I get a permeate purity alarm on my ELGA Laboratory Equipment?
    • A
      Alexandria AnthonyAug 31, 2025
      If you receive a permeate purity alarm, first press mute to silence the alarm. Then, confirm the displayed purity is suitable for the application. If issues persist, adjust the alarm setpoint or contact your local service provider.
  • W
    William BrownSep 3, 2025
    What to do if there is high loop pressure in my ELGA MEDICA Pro 60 Laboratory Equipment?
    • M
      Mary GreeneSep 3, 2025
      If there is high loop pressure, first press the mute button to silence the alarm. Then, remove the power. If the issue continues, contact Technical Support for further assistance.
  • M
    Megan CrawfordSep 6, 2025
    What to do if Protek L1 is incorrectly installed in my ELGA MEDICA Pro 60 Laboratory Equipment?
    • L
      Lisa GreerSep 6, 2025
      If the Protek L1 is incorrectly installed, first press the mute button to silence the alarm. Then, confirm that the Protek is correctly installed. Finally, power the system off and on to reset.
  • L
    Lisa BurtonSep 9, 2025
    What should I do if the Medpure L1 is incorrectly installed in my ELGA MEDICA Pro 60?
    • H
      hmeadowsSep 9, 2025
      If the Medpure L1 is incorrectly installed, first press the mute button to silence the alarm. Then, confirm that the Medpure L1 is correctly installed. Finally, power the system off and on to reset.
  • C
    Christopher HawkinsSep 12, 2025
    What to do when the UV lamp fails in my ELGA MEDICA Pro 60?
    • K
      Kenneth ChapmanSep 12, 2025
      If the UV lamp fails, first press the mute button to silence the alarm. Then, change the lamp at the next available opportunity.

Summary

INTRODUCTION

General Health and Safety

Crucial safety warnings and cautions for operating the equipment.

INSTALLATION AND COMMISSIONING

Installation Procedure

Guidance on the physical installation of the unit.

Commissioning Process

Steps to set up and configure the unit after installation.

CONTROLS

Consumable Replacement Reminders

How to set up reminders for replacing consumables.

MAINTENANCE

Replacing Composite Vent Filter

Procedure for replacing the Composite Vent Filter.

SANITIZATION

RO Sanitization Procedure

Detailed steps for sanitizing the RO module using Minncare.

EMERGENCY BYPASS

KEY TO CONTROL PANEL

TROUBLE SHOOTING

Alarm Summary and Actions

Identifies common alarms and recommended actions for troubleshooting.

TECHNICAL SPECIFICATION

ELGA MEDICA Pro 60 Specifications

General IconGeneral
Product Water Flow RateUp to 60 L/h
Endotoxin Specification<0.001 EU/mL
Resistivity18.2 MΩ·cm
Total Organic Carbon (TOC)<5 ppb
Particles >0.2 μm<1/mL
InstallationBench-top
TypeWater Purification System
Maximum Output Flow Rate60 L/h
Feed Water RequirementsPotable tap water
DisplayLCD
Bacterial Specification<0.1 CFU/mL
Output Water QualityType I
Weight45 kg
Power Supply100-240 V AC, 50-60 Hz
ApplicationsClinical analyzers, cell culture, HPLC