What to do if the ELGA CENTRA S200 US Laboratory Equipment shows alarm and flashing Resistivity value?
- JJenna WallAug 20, 2025
First, mute the alarm. Then, verify that the alarm set value is correct, referring to Section 5.2 - System preferences. Ensure the cylinder is correctly placed in line and not isolated from the distribution loop. If necessary, replace the ion-exchange cylinder, following steps 9 and 10 in Section 5 - Installation / replacement of ion-exchange cylinder.