CENTRA S200/R200 - US
Operator Manual

Questions and Answers

  • M
    michael02Aug 20, 2025
    What to do if the ELGA CENTRA-R200HFV US shows alarm and flashing Resistivity value?
    • J
      Joseph RichardsAug 20, 2025
      First, mute the alarm. Then, verify that the alarm set value is correct, referring to Section 5.2 - System preferences. Ensure the cylinder is correctly placed in line and not isolated from the distribution loop. If necessary, replace the ion-exchange cylinder, following steps 9 and 10 in Section 5 - Installation / replacement of ion-exchange cylinder.
  • R
    Richard VegaAug 22, 2025
    What to do if the ELGA CENTRA-R200HFV US Laboratory Equipment shows a high water temperature alarm?
    • M
      Michael BrownAug 22, 2025
      First, check that the correct alarm point is set, referring to section 5, steps 9 and 10 - Alarm Settings. Verify that the feedwater temperature hasn't suddenly increased. Dispense some water from the distribution loop to allow colder water to enter the unit. Recheck the feedwater temperature.
  • T
    Tristan MerrittAug 25, 2025
    What to do if the ELGA CENTRA-R200HFV US Laboratory Equipment shows a reservoir low level alarm?
    • G
      Garrett AlexanderAug 25, 2025
      Verify that the RO system is running. Note that RO production is affected by water temperature - See RO performance, section 3.4 Flowrate vs. temperature. Check if the water demand matches the water production.
  • J
    jclarkeAug 26, 2025
    Why is the ELGA Laboratory Equipment unit noise level greater than specified?
    • T
      Thomas CabreraAug 26, 2025
      Open the front door and secure the pipework to prevent vibration. Check that air isn't being drawn into the recirculation loop.
  • R
    Roger FisherAug 28, 2025
    What to do if the ELGA CENTRA-R200HFV US pumps are not running but power is on?
    • D
      Dennis ScottAug 28, 2025
      Check the thermal overloads in the electrical enclosure and reset them by pressing.
  • J
    Jason ChangAug 29, 2025
    What to do if the ELGA CENTRA-R200HFV US Laboratory Equipment shows a low feed pressure alarm?
    • J
      James JenningsAug 29, 2025
      Mute the alarm and check the feed pressure into the unit, ensuring it is greater than 2 bar at 20l/min. Press the restart button for the RO.
  • T
    Travis DuncanAug 31, 2025
    What to do if the ELGA Laboratory Equipment shows a pre-treatment filter change reminder alarm?
    • J
      Jose SalinasAug 31, 2025
      Mute the alarm and replace the filter. See section 9.2 - Replacing the pre-treatment filter.
  • T
    Trevor WhiteSep 2, 2025
    What to do if the ELGA CENTRA-R200HFV US shows a 0.2µm in line filter change reminder alarm?
    • W
      williambrandtSep 2, 2025
      Mute the alarm and replace the filter. See section 9.3 - Replacing the 0.2µm in line filter.
  • B
    Bonnie MurraySep 4, 2025
    What to do if the ELGA CENTRA-R200HFV US Laboratory Equipment shows a sanitization reminder alarm?
    • A
      Amanda RuizSep 4, 2025
      Mute the alarm and initiate the Sanitization Procedure. See Section 10 - Sanitization Procedure.
  • P
    Pamela ReynoldsSep 6, 2025
    What to do if the ELGA Laboratory Equipment shows a reservoir level disconnect fault alarm?
    • C
      craigmanuelSep 6, 2025
      Mute the alarm. Make sure the control lead from the reservoir is properly connected.

Summary

Health and Safety Precautions

General Safety Precautions

Essential safety warnings and cautions for operating and maintaining the CENTRA unit.

Electrical and Pressure Safety

Guidelines for safe electrical connections and managing system pressure.

Chemical Handling and COSHH

Safety information for sanitization chemicals and control of hazardous substances.

Product and Process Description

Installation Instructions

Unpacking and Environment Setup

Steps for unpacking the unit and preparing the installation environment.

Positioning and Hydraulic Connections

Guidance on safely positioning the unit and making hydraulic connections.

Electrical Supply Connection

Detailed steps for safely connecting the unit to the electrical supply.

Commissioning and Operation

Maintenance and Monitoring

Sanitization Procedures

Troubleshooting and Alarms

Problem Diagnosis and Solutions

Troubleshooting guide for common issues and error messages.

Alarm Condition Reference

Explanation of alarm screens, faults, and their possible causes.

Replacement Timer Reference

Guide to setting and resetting timers for component replacements.

ELGA CENTRA-R200HFV US Specifications

General IconGeneral
TypeWater Purification System
ModelCENTRA-R 200HFV US
TOC< 5 ppb
Particles >0.2 μm< 1/mL
Endotoxins< 0.001 EU/mL
Product Water QualityType 1 ultrapure water
RO Permeate Flow RateUp to 200 L/h
Resistivity18.2 MΩ-cm at 25°C
Power Supply200-240V, 50/60 Hz
Feed Water RequirementsPotable tap water