PURELAB flex 3/4 Operators Manual ELGA
Page 16 PURELAB flex 3/4 MANU39265 Version 01 - 09/10
Dimensions
Width 236mm, Depth 470mm, Height minimum 900mm,
Height maximum 1020mm.
Model
(
Ultra Pure Water (Type I)
Directly from tap water)
(
Ultra Pure Water (Type I) from
RO (Type III) feed water)
ELECTRICAL REQUIREMENTS
Mains input 100-240V ac, 50-60 Hz
System control voltage
(not including pumps and UV)
24 V dc
Power consumption (peak demand)
100 VA
PRODUCT WATER SPECIFICATION
Model
(
Ultra Pure Water (Type I)
directly from tap water)
(Ult
ra Pure Water (Type I) from
RO (Type III) feed water)
Volume of Ultra Pure (Type 1) Water
used per working day.
Up to 10 L/day
Ultra Pure delivery flow rate from
Dispense Handset
Up to 2 L/min
Inorganics (Resistivity @25°C) 18.2MΩ-cm
Organics (TOC) – typically <5ppb
1
Bacteria – Typical <1 CFU/ml (when fitted with POU filter)
Bacteria – Typical <1 CFU/10ml (when fitted with Biofilter)
Endotoxin <0.001 EU/ml (when fitted with Biofilter)
DNase <20 pg/ml (when fitted with Biofilter)
RNase 0.002 ng/ml (when fitted with Biofilter)
Particles 0.2µm filtration (when fitted with POU filter)
pH Effectively neutral
Reverse Osmosis (Type III) Make-
up flow rate
Up to 10 L/hr
Volume of Pure (Type III) Water
used per working day.
(Taken directly from the internal
<10 L/day
1
Dependent on feed water
As part of our policy of continual improvement we reserve the right to alter the