6 Emergency lid release
To open the lid of centrifuge in case of mains failure or lid lock damage, disconnect
the device from mains power supply, wait until the rotor comes to a complete stop then
lift the centrifuge and insert the included emergency lid opening key in to the groove and
move the lever to the side (as it is shown on the figure 2). The lever is located on the
bottom of centrifuge in the groove near the control panel.
Fig.2
ATTENTION! NEVER open the lid in emergency procedure when rotor is still turning!