3 Installation
Unpack the centrifuge. Please keep the box and packaging materials for transportation
or storing.
If the machine was stored at a temperature below 0°C, make sure it stays at room
temperature for at least two hours before turning on.
Place the centrifuge onto a level, horizontal surface.
Examine the power plug, power cord and appearance of centrifuge for damage.
Connect the power cord to the centrifuge, insert the plug in to the socket and press the
ON switch at (I) symbol. Display will light up on the control panel.
Open the lid by pressing the OPEN button.
Check the rotor, make sure that rotor nut is securely fastened. Try rotating it by hand.
There should be no noise or obstacles for smooth run.
Check the adapters. All the adapters must be equally positioned inside the brackets.
Check the surrounding of device. Safety distance of 30 cm must be observed around
the centrifuge during operation.
If no other damage or malfunction were determined, the centrifuge is ready to run.
ATTENTION! If there is apparent damages or malfunctions, DO NOT TURN ON the
centrifuge without consulting a specialist.
ATTENTION! To finish the unpacking of centrifuge, pull the emergency opening lever to
the side (as shown on the figure on page 12) and remove all packaging materials from the
rotor bucket.