EasyManua.ls Logo

Elmi CM-7S Plus - Page 18

Elmi CM-7S Plus
25 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17
Rotor
Description
ROTOR 8M.07
Universal rotor for 4 anodized aluminum adapters (flat bottom), also
suitable for platelet-rich plasma (PRP) preparation.
Maximal volume of applied test tubes: 50 ml.
Maximal size of applied test-tubes (D x L): 30 x 130 mm.
Top speed: 4500 rpm.
Total test tube imbalance limit: 5 g
Autoclave sterilization at a temperature up to 121 °С (20
minutes) is allowed.
ROTOR 8M.08
Universal rotor for 6 anodized aluminum adapters (flat bottom), also
suitable for platelet-rich plasma (PRP) preparation.
Maximal volume of applied test tubes: 50 ml.
Maximal size of applied test-tubes (D x L): 30 x 120 mm.
Top speed: 4500 rpm.
Total test tube imbalance limit: 5 g.
Autoclave sterilization at a temperature up to 121 °С (20
minutes) is allowed.
ROTOR 8M.09
Universal rotor for 2 anodized aluminum adapters
(cone bottom) and 4 stainless steel adapters.
Maximal size of applied test-tubes (D x L):
30 x 130 mm - 2 pcs.
17.5 x 120 mm - 4 pcs.
It is possible to order inserts in 50 ml adapters (diameter 18 and 22
mm) for the test tubes (DxL) 17.5 x 115 mm/20 x 115 mm.
Top speed: 4500 rpm.
Total test tube imbalance limit: 3g.
Autoclave sterilization at a temperature up to 121 °С (20
minutes) is allowed.
ROTOR 8M.10
Universal rotor for 2 anodized aluminum adapters (flat bottom)
and 4 stainless steel adapters, also suitable for platelet-rich
plasma (PRP) preparation.
Maximal size of applied test-tubes (D x L):
37 x 125 mm - 2 pcs.
17.5 x 120 mm - 4 pcs.
Top speed: 4500 rpm.
Total test tube imbalance limit: 3 g.
Autoclave sterilization at a temperature up to 121 °С (20
minutes) is allowed.
ROTOR 8М.18
Universal rotor for 2 anodized aluminum (cone bottom) and 2
stainless steel adapters.
Maximal size of applied test-tubes (DxL):
30 x 130 mm 2 pcs.
17,5 x 131 mm 2 pcs.
Top speed: 4500 rpm.
Total test tube imbalance limit: 3 g.
Autoclave sterilization at a temperature up to 121 °С (20
minutes) is allowed.

Related product manuals