18
Cl Calibration.
Cl calibration procedure involves probe’s selection, Zero and 2nd Point calibration.
This procedure assumes that instrument is correctly congured and a working Chlorine probe connected and installed
on system. Measurement must be performed using plant water.
Otherwise unattended results may occurr.
Select Probe.
Once into “Calibration Cl” menu move on “Select Probe” then press wheel to enter into probe list. According to system
choose most suitable probe by moving the wheel. Then press to conrm. Note: SCLxx probes series don’t require
ZERO calibration and select probe.
g. 14
Probe Scale (mg/l)
ECL 1/2 2.000mg/l Cl2
ECL 1/5 5.00mg/l Cl2
ECL 1/20 20.00mg/l Cl2
ECL 1/200 200.0mg/l Cl2
ECL 2/2 2.000mg/l Cl2O2
ECL 2/20 20.00mg/l Cl2O2
ECL 3/2 2.000mg/l Cl2
ECL 3/10 10.00mg/l Cl2
ECL4,5,6,7 10.00mg/l Cl2 or Br
ECL 8/2 2.000 mg/l Cl2Tot
ECL 8/20 20.00 mg/l Cl2Tot
ECL 9/200 200.0mg/l H2O2
ECL9/2000 2000mg/l H2O2
ECL10/1 0.5mg/l O3
ECL 10/10 10.00mg/l O3
ECL11/200 200.0mg/l peracetic acid
ECL11/2000 2000mg/l peracetic acid
ECL 13 60.00mg/l O2
ECL 17/10 10.00mg/l Cl2O2
ECL 18/10 10.00 mg/l Cl2