EasyManua.ls Logo

Emec MAX5 - Chlorine Zero Calibration; Chlorine 2 nd Point Calibration

Emec MAX5
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
Calib Zero.
Once into “Calibration Cl” menu move on “Calib Zero” then press wheel to enter into calibration mode. For a correct
system calibration proceed as follows.
- install an “activated carbon lter” prior to probe’s holder.
- let system water ow into probe holder for about 30 minutes.
- press wheel (cursor must be on “OK”).
- remove “activated carbon lter”.
Calib 2nd point.
Once into “Calibration Cl” menu move on “Calib 2nd Pt” then press wheel to enter into calibration mode. For a correct
system calibration use a Photometer or a DPD device to read chlorine on system. Enter value using the wheel the move
cursor on “OK” then press wheel.
End procedure by moving cursor on “Exit” from “Calibration Cl” main menu and press it. If an error occurred during ca-
libration procedure then the instrument will show an error message and will ask to proceed to a new calibration, cancel
current operation or restore default settings.
Carbon Filter System
Photometer