Quick Installation Guide
00825-0100-4001, Rev BA
November 2002
Model 3051
STEP 5: CONFIGURE THE TRANSMITTER
NOTE:
A check (⻫) indicates the basic configuration parameters. At minimum,
these parameters should be verified as part of the configuration and
startup procedure.
Table 1. The Model 275 HART Communicator Fast Key Sequence
Function Fast Key Sequence
Alarm and Saturation Levels 1, 4, 2, 7
Analog Output Alarm Type 1, 4, 3, 2, 4
Burst Mode Control 1, 4, 3, 3, 3
Burst Operation 1, 4, 3, 3, 3
Custom Meter Configuration 1, 3, 7, 2
Custom Meter Value 1, 4, 3, 4, 3
⻫ Damping 1, 3, 6
Date 1, 3, 4, 1
Descriptor 1, 3, 4, 2
Digital To Analog Trim (4-20 mA Output) 1, 2, 3, 2, 1
Disable Local Span/Zero Adjustment 1, 4, 4, 1, 7
Field Device Information 1, 4, 4, 1
Full Trim 1, 2, 3, 3
Keypad Input – Rerange 1, 2, 3, 1, 1
Local Zero and Span Control 1, 4, 4, 1, 7
Loop Test 1, 2, 2
Lower Sensor Trim 1, 2, 3, 3, 2
Message 1, 3, 4, 3
Meter Options 1, 4, 3, 4
Number of Requested Preambles 1, 4, 3, 3, 2
Poll Address 1, 4, 3, 3, 1
Poll a Multidropped Transmitter Left Arrow, 4, 1, 1
⻫ Range Values 1, 3, 3
Rerange 1, 2, 3, 1
4001_RevBA.fm Page 8 Monday, November 25, 2002 2:27 PM