140 Micro Motion Series 1000 and Series 2000 Transmitters
Troubleshooting
A108
(5)
Event #1 Triggered Event 1 Triggered Be advised of alarm condition.
If you believe the event has been triggered
erroneously, verify the Event 1 settings. See
Section 8.11.
A109
(5)
Event #2 Triggered Event 2 Triggered Be advised of alarm condition.
If you believe the event has been triggered
erroneously, verify the Event 2 settings. See
Section 8.11.
A110 Frequency Output
Saturated
Frequency Output
Saturated
See Section 12.18.
A111 Frequency Output Fixed Frequency Output Fixed Exit frequency output loop test.
A112
(6)
Software upgrade
recommended
S/W Upgrade
Recommended
Contact Micro Motion to get a Series 1000/2000
transmitter software upgrade. See Section 12.3.
Note that the device is still functional.
A113 Secondary mA Output
Saturated
Secondary mA Output
Saturated
See Section 12.18.
A114 Secondary mA Output
Fixed
Secondary mA Output
Fixed
Exit mA output loop test. See Section 5.3.
Exit mA output trim. See Section 5.4.
Check to see if the output has been fixed via digital
communication.
A115 External Input Error External Input Error HART polling connection to external device has
failed. Ensure that external device is available:
• Verify device operation.
• Verify wiring.
Verify polling configuration. See Section 9.4.
A116 API Temperature
Out-of-Limits
API: Temperature
Outside Standard
Range
Verify process.
Verify API reference table and temperature
configuration. See Section 8.6.
A117 API Density
Out-of-Limits
API: Density Outside
Standard Range
Verify process.
Verify API reference table and density configuration.
See Section 8.6.
A118 Discrete Output 1 Fixed Discrete Output 1 Fixed Exit discrete output loop test. See Section 5.3.
A119 Discrete Output 2 Fixed Discrete Output 2 Fixed Exit discrete output loop test. See Section 5.3.
A120 ED: Unable to Fit Curve
Data
ED: Unable to Fit Curve
Data
Verify enhanced density configuration.
A121 ED: Extrapolation Alarm ED: Extrapolation Alarm Verify process temperature.
Verify process density.
Verify enhanced density configuration.
A131
(3)
Meter Verification Info
Alarm
Meter
Verification/Outputs at
Last Value
Meter verification in progress, with outputs set to
last measured value. Allow the procedure to
complete. If desired, abort the procedure and restart
with outputs set to fault.
A131
(4)
Meter Verification in
Progress
Meter Verification In
Progress
Meter verification in progress, with outputs set to
continue reporting process data. Allow the
procedure to complete.
A132
(2)
Simulation Mode Active Simulation Mode Active Disable simulation mode. See Section 12.10.
Table 12-5 Status alarms and remedies continued
Alarm
code Communicator ProLink II software Possible remedy

Table of Contents

Other manuals for Emerson Micro Motion 2500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Micro Motion 2500 and is the answer not in the manual?

Emerson Micro Motion 2500 Specifications

General IconGeneral
BrandEmerson
ModelMicro Motion 2500
CategoryTransmitter
LanguageEnglish

Related product manuals