7.2 HART/analog single loop wiring
HART/analog single loop wiringFigure 7-2:
A. 820 Ω maximum loop resistance
B. HART-compatible host or controller
Note
For HART communications:
• 600 Ω maximum loop resistance
• 250 Ω minimum loop resistance
I/O wiring for Model 1700 and Model 2700 transmitters with analog outputs
78 Micro Motion
®
Model 1700 and 2700