Integratethemeterwiththecontrolsystem
OptionsforHARTvariables
Table2-13OptionsforHARTvariables
ProcessvariablePVSVTVQV
Massowrate
üüüü
Volumeowrate
üüüü
Temperature
üüü
Density
üüü
Drivegain
üüü
Masstotal
ü
Volumetotal
ü
Massinventory
ü
Volumeinventory
ü
Tubefrequency
ü
Metertemperature
ü
LPOamplitude
ü
RPOamplitude
ü
Boardtemperature
ü
Externalpressure
(16)
üüü
Externaltemperature
(16)
üüü
Gasstandardvolumeowrate
(16)
üüüü
Gasstandardvolumetotal
(16)
ü
Gasstandardvolumeinventory
(16)
ü
Livezero
ü
Temperature-corrected(standard)volumeowrate
(17)
üüüü
Temperature-corrected(standard)volumetotal
(17)
ü
Temperature-corrected(standard)volume
inventory
(17)
ü
Averagetemperature
(17)
üüü
Averagedensity
(17)
üüü
CTL
(17)
ü
Densityatreferencetemperature
(18)
üüü
Specicgravity
(18)
üüü
Standardvolumeowrate
(18)
üüüü
Standardvolumetotal
(18)
ü
Standardvolumeinventory
(18)
ü
(16)Requirestransmittersoftwarev5.0orlater.
(17)Availableonlyifthepetroleummeasurementapplicationisenabledonyourtransmitter.
(18)Availableonlyiftheconcentrationmeasurementapplicationisenabledonyourtransmitter.