2 Mounting and sensor wiring
2.1 Mounting and sensor wiring for integral-mount
transmitters
There are no separate mounting requirements for integral transmitters, and there is no
need to connect wiring between the transmitter and the sensor.
2.2 Mount the transmitter to a wall or instrument
pole
There are two options available for mounting the transmitter:
• Mount the transmitter to a wall or flat surface.
• Mount the transmitter to an instrument pole.
Prerequisites
• If you are mounting the transmitter to a wall or flat surface:
— Ensure that the surface is flat and rigid and that it does not vibrate or move
excessively.
— Confirm that you have the necessary tools and the mounting kit shipped with the
transmitter.
— Confirm that the mounting surface, method, and surface structure ensures
sufficient strength to secure the transmitter (for example, when mounting to drywall
use a toggle type drywall anchor).
• If you are mounting the transmitter to an instrument pole:
— Ensure that the instrument pole extends at least 12in (305mm) from a rigid base
and is no more than 2in (51mm) in diameter.
— Confirm that you have the necessary tools and the instrument-pole mounting kit
shipped with the transmitter.
Installation Manual Mounting and sensor wiring
MMI-20058013 March2024
Installation Manual 11