Figure 2-13: Rotation of the sensor wiring junction box
3. Gently set the junction box into its new position, confirming that the position is
locked.
4. Replace the clamp in its original position and tighten the cap screw. Tighten the
screw to 28in lbf (3.16N m)– 32in lbf (3.62N m).
Figure 2-14: Re-attachment of the clamp
Mounting and sensor wiring Installation Manual
March2024 MMI-20058013
22 Micro Motion 4200 2-Wire Transmitter