3.2.3 Prepare a cable with user-supplied cable glands
1. Remove the core processor cover using a flat-blade screw driver.
2. Pass the wires through the gland nut and gland body.
A. Gland body
B. Gland nut
3. Terminate the RS-485 shield and drain wires to the housing internal grounding
screw.
4. Assemble the gland according to vendor instructions.
3.2.4 Prepare a cable with Micro Motion-supplied cable glands
1. Remove the core processor cover using a flat-blade screw driver.
2. Pass the wires through the gland nut and clamping insert.
A. Gland nut
B. Clamping insert
3. Strip the cable jacket.
Option
Description
NPT gland type Strip 4-1/2 inch (115 mm)
M20 gland type Strip 4-1/4 inch (108 mm)
4. Remove the clear wrap and filler material.
5. Strip most of the shielding.
Option
Description
NPT gland type Strip all but 3/4 inch (19 mm)
M20 gland type Strip all but 1/2 inch (12 mm)
6. Wrap the drain wires twice around the shield and cut off the excess drain wires.
Transmitter power and I/O wiring