9 Hazardous location certifications
9.1 US and Canadian Certifications
E5
or
E6
Explosion-Proof for Class I, Division 1, Groups B, C, and D
Dust ignition-proof for Class II, III Division 1, Groups E, F, G.
Ex db [ia] IIC T6...T1 Gb
Class I, Zone 1, AEx db [ia] IIC T6...T1 Gb (-50 °C ≤ Ta ≤ 70 °C)
Thermal data:
Ambient temperature
(°C)
Process temperature
(°C)
T-Class sensor (°C)
–50 to +70 –200 to +75 T6
–50 to +70 –200 to +95 T5
–50 to +70 –200 to +130 T4
–50 to +70 –200 to +195 T3
–50 to +70 –200 to +290 T2
–50 to +70 –200 to +427 T1
Factory Sealed; Single Seal
Enclosure Type 4X
Install per drawing 08800-0112;
I5 or
I6
Intrinsically safe for use in Class I, II, III Division 1, Groups A, B, C, D, E,
F, G
Non-incendive for Class I, Division 2, Groups A, B, C and D
Class I, Zone 0, AEx ia IIC T4 Ga
Temperature Code T4 (-50 °C ≤ Ta ≤ 70 °C) 4-20 mA HART
Temperature Code T4 (-50 °C ≤ Ta ≤ 60 °C) Fieldbus
Single Seal; Enclosure Type 4X
Install per drawing 08800-0112
IE or
IF
FISCO for Class I, Division 1, Groups A, B, C, and D
FNICO for Class I Division 2, Groups A, B, C, and D
Class I, Zone 0, AEx ia IIC T4 Ga
Temperature Code T4 (-50 °C ≤ Ta ≤ 60 °C)
Single Seal; Enclosure Type 4X
Install per drawing 08800-0112