4.4  Sand noise calibration
It is recommended to complete sand noise calibration for the device
to ensure that it has the highest accuracy possible. During this,
sand is injected in known quantities and velocities. This will generate
calibration curves which will allow the device to detect sand and
quantify it.
Note
Additional equipment (sand injector) will be required to complete this
phase.
Additional information can be given by contacting your local service
team representative.
The process details the device level sand calibration.
For this process, a minimum of six data collections are required:
• Three variable flow velocities with a fixed sand rate
— Flow rates are suggested to be minimum, maximum, and
median of expected operational velocities.
• Three fixed flow velocity with variable sand rates
Connection to the device is the same as described in Connecting to
the device and commissioning homescreen.
May2024 Quick Start Guide