Field cable is not supplied with the product as standard. Field cable is
orderable and supplied separately to the device.
1.2 Required tools and equipment for installation
This section lists the required tools and equipment for physical
installation, configuration and commissioning of the Rosemount
SAM42 Acoustic Particle Monitor.
1.2.1 Configuration and commissioning
The Rosemount SAM42 Acoustic Particle Monitor must be configured
before physical installation. Direct configuration must be performed
using a Windows
™
laptop or tablet with the SAM42 Commissioning
app installed.
• Computer or tablet running Windows
• USB to RS 485 Converter
• SAM42 Commissioning app
Note
If using a tablet or laptop in the hazardous area, ensure that the
correct permits are applied and granted before carrying out any of
the work.
SAM42 Commissioning App is available for download on Emerson
Software Application Portal.
USB to RS 485 Converter is included in the installation kit that is
available for purchase upon request. Typically, one installation kit per
site should suffice.
1.2.2 Physical installation
The tools required for mounting the device in the field are the
following:
• Tinsnips
• Screwdriver, flathead
• Spanner, 13 mm
• Spanner, custom to fit device lid, nut height gauge also built in
• Socket, 8 mm, ¼-in. drive
• Torque wrench ¼-in. drive, 2.5 N-m to 15 N-m
• Allen key, 3 mm
• Sandpaper (grade 60-100)/ wire brush for sensitivity testing
• Flat file, 250 mm
May2024 Quick Start Guide
Quick Start Guide 7