Cleanfit CPA871 Mounting
Endress+Hauser 29
Connecting the cleaning unit
As-delivered state, PN16 with dummy plug and pressure retarder.
1. Replace the dummy plug with the rinse connection inlet.→  28
2.
A0043236
Mount the connection for the rinse medium supply line on the rinse connection inlet.
3.
A0043237
Connect the rinse connection outlet to a drain line.
Complete seal PN8 and PN16
Complete seal with pressure retarder and ball valve
The complete seal with the pressure retarder and ball valve is only provided in the PN16
version. The pressure retarder must be mounted for this purpose. The ball valve can be
ordered as an accessory (sealing kit).
The sealing kit only provided in connection with the rinse connection inlet G1/4". Version only
available in stainless steel.
A0043406
Not suitable for media that tend to cake, form deposits or sediment, or contain solids.
Screw the ball valve onto the rinse connection outlet of the pressure retarder.
Complete seal with extension and ball valve
The complete seal with the extension and ball valve is only provided in the PN8 version. The
extension and ball valve can be ordered as an accessory. The extension is only necessary for
the flange process connection. The ball valve and extension can be ordered as an accessory
(sealing kit).
Not suitable for media that tend to cake, form deposits or sediment, or contain solids.