Cleanfit CPA871 Mounting
Endress+Hauser 37
2. Remove the yellow dummy plug (item 3).
3. Use the open-ended wrench (item 1) to screw in the sensor (item 4) in place of the
dummy plug and hand-tighten (3 Nm (2.2 lbf ft)).
4. Secure the open-ended wrench back in the cover.
5. Mount the cover on the assembly. When doing so, guide the measuring cable through
the cable grommet (top of cover).
Always mount the cover before you move the assembly to the measuring position. The
cover cannot be removed in the measuring position and therefore prevents the removal
of the sensor.