Cleanfit CPA871 Maintenance
Endress+Hauser 65
Assembly
In the immersion chamber version, the inlet and outlet of the service chamber are fixed.
When assembling the immersion tube, please ensure that the leakage hole (pos.11), the
service chamber outlet (pos. 12) and the immersion chamber (pos. 16) are all in one line.
Fit the components of the immersion chamber together.
A0030368
46 Assembling the immersion chamber
8
9
10
11
12
13
15
16
Immersion chamber - top part
Immersion chamber - middle part
Immersion chamber - bottom part
Leakage hole
Rinse connection - outlet of service chamber
Positioning pin
Positioning pin
Position marking
1. Fit the bottom part (pos. 10) and middle part (pos. 9) of the immersion chamber
together. Ensure correct orientation!
2. Screw both parts securely together using the three securing screws (pos. 11).
3. Attach the top part (pos. 8) of the immersion chamber.