Maintenance Cleanfit CPA871
66 Endress+Hauser
A0030347
47 Assembling the support housing and immersion chamber
4 Support housing
13 Positioning pin
14 Positioning groove
4. Place the support housing (pos. 7) on a level surface.
The positioning groove (pos. 14) is visible from above.
5. Place the immersion chamber on the support housing.
6. Slide the service chamber into the support housing.
7. Place the positioning pin (pos. 13) over the corresponding groove.
8. Push the service chamber into the groove.
9. Mount the process connection on the support housing.