EasyManua.ls Logo

Endress+Hauser CM14-AAK - Page 30

Endress+Hauser CM14-AAK
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting CM14
30 Endress+Hauser
Diagnostics
code
Message text Description
F846 Param error
Faulty parameter checksum
Possible cause:
Firmware update
Remedy:
Reset parameters to factory defaults
F847 Couldn't save param Parameters could not be saved
F848 Calib AO1 Faulty calibration values for analog output 1
F849 Calib AO2 Faulty calibration values for analog output 2
F904 Process check
Process check system alarm.
No change in measurement signal for a long time.
Possible reasons
Sensor dirty or in air
No sensor inflow
Sensor defective
Software error
Remedy:
Check measuring chain
Inspect sensor
Carry out software restart
Diagnostics
code
Message text Description
C107 Calib. active
Sensor calibration is active.
Remedy:
Wait for calibration
C154 No calib. data
Sensor data.
No calibration data present, factory settings will be used.
Remedy:
Check calibration information of the sensor
Calibrate cell constant
C850 Simu AO1 Simulation of analog output 1 is active
C851 Simu AO2 Simulation of analog output 2 is active
C853 Download act. Parameter transmission is active

Related product manuals