CM14 Troubleshooting
Endress+Hauser 31
Diagnostics
code
Message text Description
S844 Process value
Measured value outside the specified range.
Measured value outside the specified range.
Possible reasons:
Sensor in air
Air cushion in the assembly
Incorrect sensor inflow
Sensor defective
Remedy:
Increase process value
Check measuring chain
Change sensor type
S910 Limit switch Limit switch energized
Diagnostics
code
Message text Description
M126 Sensor check
Check sensor.
Bad electrode status.
Possible reasons:
Glass membrane blocked or dry
Junction blocked
Remedy:
Clean sensor and regenerate
Replace sensor
M500 Not stable
Sensor calibration aborted.
Main measured value fluctuating.
Possible reasons:
Sensor too old
Sensor dry intermittently
Buffer value not constant
Remedy:
Check sensor, replace if necessary
Check buffer