CM14 Troubleshooting
Endress+Hauser 31
Diagnostics
code
Message text Description
S844 Process value
Measured value outside the specified range.
Measured value outside the specified range.
Possible reasons:
• Sensor in air
• Air cushion in the assembly
• Incorrect sensor inflow
• Sensor defective
Remedy:
• Increase process value
• Check measuring chain
• Change sensor type
S910 Limit switch Limit switch energized
Diagnostics
code
Message text Description
M126 Sensor check
Check sensor.
Bad electrode status.
Possible reasons:
• Glass membrane blocked or dry
• Junction blocked
Remedy:
• Clean sensor and regenerate
• Replace sensor
M500 Not stable
Sensor calibration aborted.
Main measured value fluctuating.
Possible reasons:
• Sensor too old
• Sensor dry intermittently
• Buffer value not constant
Remedy:
• Check sensor, replace if necessary
• Check buffer