Description of device parameters Deltabar FMD71, FMD72
28
Navigation Expert → System → Operator Code (021)
Write permission Operators/Service engineers/Expert
Description Use this function to enter a code to lock or unlock operation.
Entry
To lock: Enter a number ¹ the release code (value range: 1 to 9999).
To unlock: Enter the release code.
Note The release code is "0" in the order configuration. Another release code can be defined in
the "Code Definition" parameter. If the user has forgotten the release code, the release code
can be visible by entering the number "5864".
Factory setting 0
Device tag (022)
Navigation Expert → System → Instrument info → Device tag (022)
Write permission Operators/Service engineers/Expert
Description Enter the device tag, e.g. TAG number (max. 32 alphanumeric characters).
Factory setting No entry or according to order specifications
Code Definition (023)
Navigation Expert → System → Code Definition (023)
Write permission Operators/Service engineers/Expert
Description Use this function to enter a release code with which the device can be unlocked.
Options A number from 0 to 9999
Factory setting 0
Level Selection (024)
Navigation Expert → Measurement → Level → Level Selection (024)
Write permission Operators/Service engineers/Expert
Description Select the method for calculating the level