Deltabar FMD71, FMD72 Description of device parameters
27
Description
Enter damping time (time constant t) ("Damping" DIP switch set to "on")
Display damping time (time constant t) ("Damping" DIP switch set to "off").
The damping affects the speed at which the measured value reacts to changes in pressure.
Input range 0.0 to 999.0 s
Factory setting 2.0 sec. or according to order specifications
Level before lin. (019)
Navigation Expert → Measurement → Level → Level Before Lin. (019)
Write permission No write permissions. Parameter is read only.
Description Displays the level value prior to linearization.
Meas.Diff.Press. (020)
Navigation Expert → Measurement → Pressure → Meas.Diff.Press. (020)
Write permission No write permissions. Parameter is read only.
Description Displays the measured differential pressure after sensor trim, position adjustment and
damping.
Sensor press.
LP
Measured
pressure LP
LP
Sensor
calibration LP
Sensor trim LP
Pressure
simulation LP
Corrected
pressure
Pressure after
damping
Measured
differential
pressure
Differential
pressure +
pressure
inversion
Pos. zero
adjust (calib.
offset)
Differential
pressure
simulation
Damping P
HP
Sensor
calibration HP
Sensor trim
HP
Pressure
simulation HP
Sensor press.
HP
Measured
pressure HP
Operator Code (021)