Deltabar FMD71, FMD72 Description of device parameters
41
Description Display of Device Revision (e.g. 1)
Sensor pressure HP (109)
Navigation Expert → Measurement → Pressure → Sensor Press. HP (109)
Write permission No write permissions. Parameter is read only.
Description Displays the measured pressure before the sensor trim.
Sensor Temp. HP (110)
Navigation Expert → Measurement → Basic setup → Sensor Temp. HP (110)
Write permission No write permissions. Parameter is read only.
Description Displays the temperature currently measured in the sensor module. This can deviate from
the process temperature.
Simulation mode (112)
Navigation Expert → Diagnosis → Simulation → Simulation mode (112)
Write permission Operators/Service engineers/Expert
Description Switch on simulation and select the simulation mode. When changing the measuring
mode or the level type (Lin. Mode (037)) or when the device is restarted, any simulation
running is switched off.
Options • None
• Differential pressure, → see this table, "Sim. pressure" parameter
• Level, → see this table, "Sim. level" parameter
• Press. HP, → see this table, "Sim. Press. HP" parameter
• Press. LP, → see this table, "Sim. Press. LP" parameter
• Tank content, → see this table, Sim. tank cont." parameter
• Current, → see this table, "Sim. Current" parameter
• Alarm/warning, → see this table, "Sim. error no."
Factory setting None
Simulation
pressure LP
Measured
pressure LP
↓ ↑
LP →
Processing
Sensor LP measurement
→ +
Corrected
pressure
Simulation
differential
pressure