Description of device parameters Deltabar FMD71, FMD72
62
Hi trim measured (276)
Navigation Expert → Measurement → Sensor Trim LP → Hi Trim Measured (276)
Write permission No write permissions. Parameter is read only.
Description Displays the reference pressure present to be accepted for the upper calibration point.
Lo trim sensor (277)
Navigation Expert → Measurement → Sensor Trim LP → Lo Trim Sensor (277)
Write permission No write permissions. Parameter is read only.
Description Sensor module recalibration by entering a target pressure while simultaneously and
automatically accepting a reference pressure present for the lower calibration point.
Hi trim sensor (278)
Navigation Expert → Measurement → Sensor Trim LP → Hi Trim Sensor (278)
Write permission No write permissions. Parameter is read only.
Description Sensor module recalibration by entering a target pressure while simultaneously and
automatically accepting a reference pressure present for the upper calibration point.
Device type code (279)
Navigation Expert → Communication → HART info → Device ID (279)
Write permission No write permissions. Parameter is read only.
Description Display of the numerical ID of the device
39
Sensor pressure LP (280)
Navigation Expert → Measurement → Pressure → Sensor Press. LP (280)