5.2 Commissioning with an operating menu
5.2.1 Selecting the language, measuring mode and pressure unit
Language selection
Measuring mode selection
Parameter name Description
Language (000)
Selection
Menu path:
Main menu → Language
Select the menu language for the local display.
Options:
• English
• Another language (as selected when ordering the device)
• Possibly a third language (language of the manufacturing plant)
Factory setting:
English
Parameter name Description
Measuring mode
(005)
Selection
Menu path: Setup
→ Measuring mode
Select the measuring mode.
The operating menu is structured differently depending on the measuring mode selected.
!
Note!
If the measuring mode is changed, no conversion takes place. If necessary, the device has to
be recalibrated after the measuring mode has been changed.
Options:
• Pressure
• Level
•Flow
Factory setting:
Pressure