4.2 Basic Settings using the HART Communicator DXR 275
The FEC 12 allows calibration in % only. A remote calibration from the handheld
allows, e.g. technical units to be displayed. If a recalibration is to be carried out, then
a reset should be done first. The parameters of the electronic insert must not be
locked, see above!
Note!
When calibrating the FEC 12 electronic insert, the following sections will assume the
operating steps given below. They apply to all instructions and will, therefore, no
longer be given in this manual:
First step: Move from the
"MATRIX GROUP SEL."
menu to the next higher menu with the
arrow key.
Last step: Return to the
"Online"
menu with F3 [HOME].
Reset (factory settings)
Step Entry Cursor in display at Significance
13
SERVICE/SIMULATION
2 4
DEFAULT VALUES
312
12
Code number for reset
4 F4
[ENTER]
Confirms entry
5F2
[SEND]
Value to be transmitted
The factory settings given on reset are shown in the following table:
Menu
field
Significance Factory
setting
Values
entered
1; 2 EMPTY CALIBRATION [%] 0.0
1; 3 FULL CALIBRATION [%] 100.0
1; 4 MIN. CURRENT 4 mA OFF
1; 5 OUTPUT DAMPING [s] 1
1; 6 VALUE FOR 4 mA [%] 0.0
1; 7 VALUE FOR 20 mA [%] 100.0
1; 8 SAFETY ALARM MAX (110 %)
2; 1 LINEARISATION LINEAR
2; 2 OFFSET [pF] 349.90
2; 3 SENSITIVITY[pF/%] 16.49
4; 1 SET TAG NUMBER ’--------’
4; 6 SELECT UNIT %
Note!
Electronic insert FEC 12 (HART) 4 Calibrating