Electrical connection iTEMP TMT84
14 Endress+Hauser
Sensor input 1
RTD or resistance
transmitter, four-wire
- - - -
Thermocouple (TC),
voltage transmitter
Connecting to spring terminals
2.
3.
4.
5.
1.
2.
3.
A
B
C
A0039468
 6 Spring terminal connection, using the example of a head transmitter
Item A, solid wire:
1. Strip wire end. Min. stripping length 10 mm (0.39 in).
2. Insert the wire end into the terminal.
3. Pull the wire gently to ensure it is connected correctly. Repeat from step 1 if necessary.
Item B, fine-strand wire without ferrule:
1. Strip wire end. Min. stripping length 10 mm (0.39 in).
2. Press down on the lever opener.
3. Insert the wire end into the terminal.
4. Release lever opener.
5. Pull the wire gently to ensure it is connected correctly. Repeat from step 1 if necessary.