Endress+Hauser 17
Device condition and specifications Notes
Do the cables used meet the required specifications? Fieldbus cable,
Sensor cable,
→  13
Do the cables have adequate strain relief? --
Are the power supply and signal cables connected correctly? →  13
Are all the screw terminals well tightened and have the connections of the spring terminals
been checked?
→  14
Are all cable entries mounted, tightened and leak-tight? --
Are all housing covers installed and firmly tightened? --
Electrical connection of the fieldbus system Notes
Are all the connecting components (T-boxes, junction boxes, connectors, etc.) connected with
each other correctly?
--
Has each fieldbus segment been terminated at both ends with a bus terminator? --
Has the max. length of the fieldbus cable been observed in accordance with the fieldbus
specifications?
Has the max. length of the spurs been observed in accordance with the fieldbus specifications?
Is the fieldbus cable fully shielded and correctly grounded?
6 Operation options
6.1 Measured value display and operating elements
6.1.1 Option: Display TID10 for head transmitter
The display may also be subsequently
ordered at any time after purchasing the
transmitter, see the 'Accessories' section in
the Operating Instructions for the device.
1.
2.
3.
A0010227
 9 Attaching the display to the transmitter