Levelflex FMP51 Modbus Electrical connection
Endress+Hauser 21
6.2 Connecting the device
L
WARNING
Explosion Hazard!
Observe applicable national standards.
Comply with the specifications in the Safety Instructions (XA).
Use specified cable glands only.
Check to ensure that the power supply matches the information on the nameplate.
Switch off the power supply before connecting the device.
Connect the potential matching line to the outer ground terminal before applying the
power supply.
Required tools/accessories:
For devices with a cover lock: Allen key AF3
Wire stripper
When using stranded cables: One ferrule for every wire to be connected.
2.
3.
1.
10 (0.4)
10 (0.4)
4.
5.
5.
3 mm
A0035428
 2 Engineering unit: mm (in)
1. Loosen the screw of the securing clamp on the connection compartment cover and turn
the securing clamp 90° counterclockwise.
2. Unscrew the connection compartment cover.
3. Push the cable through the cable entry. To ensure tight sealing, do not remove the
sealing ring from the cable entry.