5. Setup → Tube diameter (only for "Tank type" = "Bypass / pipe")
Specify the diameter of the bypass or stilling well.
6. Setup → Tank level
Specify the filling level (Partially filled or Fully flooded)
7. Setup → Distance to upper connection
In bypasses: Specify the distance from the reference point R to the lower edge of the
upper outlet; in all other cases, keep the factory setting
8. Setup → DC value
Specify the dielectric constant of the upper medium
9. Setup → Empty calibration
Specify the empty distance E (distance from the reference point R to the 0% mark).
10. Setup → Full calibration
Specify the full distance F (distance from the 0% mark to the 100% mark).
11. Setup → Level
Displays the measured level L
L
.
12. Setup → Interface
Displays the interface height L
I
.
13. Setup → Distance
Displays the distance D
L
between the reference point R and the level L
L
.
14. Setup → Interface distance
Displays the distance D
I
between the reference point R and the interface L
I
.
15. Setup → Signal quality
Displays the signal quality of the analyzed level echo.
16. Setup → Mapping → Confirm distance
Compare the distance displayed with the actual value to start recording an
interference echo map.