Why the measured values for the interface and the total level are identical in Endress+Hauser Levelflex FMP51?
- BburgessmicheleSep 5, 2025
The measured values for the interface and the total level can be identical if the echo threshold for the total level is too high due to a wrong dielectric constant. To solve this, enter the correct dielectric constant (DC value) of the upper medium.