Mounting Levelflex FMP56, FMP57 FOUNDATION Fieldbus
26 Endress+Hauser
1. Unscrew the securing screw using an open-ended wrench.
2. Rotate the housing in the desired direction.
3. Tighten the securing screw (1.5 Nm for plastic housing; 2.5 Nm for aluminum or
stainless steel housing).
6.2.6 Turning the display
Opening cover
A0021430
1. Loosen the screw of the securing clamp of the electronics compartment cover using an
Allen key (3 mm) and turn the clamp 90 ° counterclockwise.
2. Unscrew cover and check lid gasket, replace if necessary.
Turning the display module
A0036401
1. Pull out the display module with a gentle rotational movement.
2. Rotate the display module to the desired position: max. 8 × 45 ° in each direction.