Liquiphant FTL31 IO-Link System integration
Endress+Hauser 25
Designation
ISDU (dec)
ISDU (hex)
Size (byte)
Data type
Access Default
value
Value range Offset/
gradient
Data
storage
Range
limits
Device search
69
0x0045
1
UInt8
r/w 0~off 0 ~ off
1 ~ on
0 / 0 No 0..1
Sensor check
70
0x0046
1
UInt8
-/w 0 / 0 No
Active switchpoints
(OU1)
64
0x0040
1
UInt8
r/w 0 ~
Density
>0.7g/cm³
0 ~ Density
>0.7g/cm³
1 ~ Density
>0.5g/cm³
2 ~ User
0..2
Reset user
switchpoints
65
0x0041
1
UIntegerT
r/w 0 ~ False 0 ~ False
1 ~ switchpoints
Ou1
0..1
Switch point value,
Output 1 (SP1/FH1)
71
0x0047
2
UInt16
r/w 88.0 0 / 1 Yes 45 to 97
Switchback point
value, Output 1 (rP1/
FL1)
72
0x0048
2
UInt16
r/w 91.0 0 / 1 Yes 45 to 97
Switching delay time,
Output 1 (dS1)
81
0x0051
2
UInt16
r/w 0.5 0 / 0.1 Yes 0.3 to 60
Switchback delay
time, Output 1 (dR1)
82
0x0052
2
UInt16
r/w 1 0 / 0.1 Yes 0.3 to 60
Output 1 (OU1)
85
0x0055
1
UInt8
r/w 0~HNO 0 ~ HNO
1 ~ HNC
2 ~ FNO
3 ~ FNC
Yes 0..3
Output 1 (OU1)
101
0x0065
1
UInt8
r/w 0~HNO 0 ~ HNO
1 ~ HNC
Yes 0..1
Active switchpoints
(OU2)
77
0x004D
1
UInt8
r/w 0 ~
Density
>0.7g/cm
0 ~ Density
>0.7g/cm³
1 ~ Density
>0.5g/cm³
2 ~ User
0..2