Technical data Liquiphant FTL31 IO-Link
52 Endress+Hauser
0 +50 +100 +150
–40
+32
+122 +212 +302
–40
–40
+70
[°C]
–40
+158
[°F]
0
+32
[°C]
[°F]
+50
+122
T
a
T
P
1
2
T
a
T
p
+90
+194
A0020869
16 Derating curve: 150 °C (302 °F)
1 I
max
: 200 mA (DC-PNP)
2 I
max
: 150 mA (DC-PNP)
Ta Ambient temperature
Tp Process temperature
17.3 Process
Note the pressure and temperature derating depending on the selected process
connection
Process temperature range –40 to +100 °C (–40 to +212 °F)
–40 to +150 °C (–40 to +302 °F)
Process pressure range max. –1 to +40 bar (–14.5 to +580 psi)
Density
>0.7 g/cm
3
(optionally available: >0.5 g/cm
3
), can be configured via IO-Link
State of aggregation Liquid
Viscosity 1 to 10 000 mPa·s dynamic viscosity
Solids contents ø < 5 mm (0.2 in)
Lateral loading capacity Lateral loading capacity of the tuning fork: max. 200 N