Mounting Liquiphant FTL51B
14 Endress+Hauser
Installing in pipes
• Flow velocities up to 5 m/s at a viscosity of 1 mm
2
/s(cSt) and density 1 g/cm
3
(SGU)
Check for correct functioning in the event of other process medium conditions.
• The flow will not be significantly impeded if the tuning fork is correctly aligned and the
marking is pointing in the direction of flow.
• The marking can be identified while the device is installed
A0034851
12 Marking and fork position
Screwing in the device
A0034852
13 Screwing in the device
• Turn by the hex bolt only, 15 to 30 Nm (11 to 22 lbf ft)
• Do not turn at the housing!