Liquistation CSF48 Electrical connection
Endress+Hauser 31
6 Electrical connection
L
WARNING
Device is live!
Incorrect connection may result in injury or death!
The electrical connection may be performed only by an electrical technician.
The electrical technician must have read and understood these Operating Instructions and
must follow the instructions contained therein.
Prior to commencing connection work, ensure that no voltage is present on any cable.
NOTICE
The device does not have a power switch
A fuse with a maximum rating of 10 A must be provided by the customer. Observe the local
regulations for installation.
Use a HBC fuse with 10 A, 250 V AC for samplers with a CSA approval
The circuit breaker must be a switch or power switch, and you must label it as the circuit
breaker for the device.
The protective ground connection must be established before all other connections. If the
protective ground is disconnected, this can be a source of danger.
A circuit breaker must be located near the device.
For 24V versions, the power supply at the voltage source must be isolated from cables
carrying low voltage (110/230V AC) by double or reinforced insulation.
Operation with non-stationary mains cable connection to sampler (optional)
NOTICE
The device does not have a power switch
The mains switch can be ordered via a TSP modification.
A fuse with a maximum rating of 10 A must be provided internally when operating with a
power cable. The fuse can be mounted under the rear cover.
The protective ground connection must be established before all other connections. If the
protective ground is disconnected, this can be a source of danger.