Liquistation CSF48 Electrical connection
Endress+Hauser 31
6 Electrical connection
L
WARNING
Device is live!
Incorrect connection may result in injury or death!
‣
The electrical connection may be performed only by an electrical technician.
‣
The electrical technician must have read and understood these Operating Instructions and
must follow the instructions contained therein.
‣
Prior to commencing connection work, ensure that no voltage is present on any cable.
NOTICE
The device does not have a power switch
‣
A fuse with a maximum rating of 10 A must be provided by the customer. Observe the local
regulations for installation.
‣
Use a HBC fuse with 10 A, 250 V AC for samplers with a CSA approval
‣
The circuit breaker must be a switch or power switch, and you must label it as the circuit
breaker for the device.
‣
The protective ground connection must be established before all other connections. If the
protective ground is disconnected, this can be a source of danger.
‣
A circuit breaker must be located near the device.
‣
For 24V versions, the power supply at the voltage source must be isolated from cables
carrying low voltage (110/230V AC) by double or reinforced insulation.
Operation with non-stationary mains cable connection to sampler (optional)
NOTICE
The device does not have a power switch
‣
The mains switch can be ordered via a TSP modification.
‣
A fuse with a maximum rating of 10 A must be provided internally when operating with a
power cable. The fuse can be mounted under the rear cover.
‣
The protective ground connection must be established before all other connections. If the
protective ground is disconnected, this can be a source of danger.