Electrical connection Liquistation CSF48
50 Endress+Hauser
6.5.3 Relays
Module 2R Module 4R
 53 Module
41
43
42
Relay 1
41
43
42
Relay 2
 54 Wiring
diagram
 55 Module
41
43
42
Relay 3
41
43
42
Relay 4
41
43
42
Relay 1
41
43
42
Relay 2
 56 Wiring
diagram
A maximum of 4 relay outputs are supported.
6.6 Connecting the supply voltage
6.6.1 Laying the cable
Lay the cables so that they are protected behind the rear panel of the device.
Cable glands (up to 8 depending on the version) are available for the cable entry.
The cable length from the foundation to the terminal connection is approx. 1.7 m (5.6 ft).
For analyzer stands, the cable length is approx. 1.8 m (5.9 ft) from the foundation.
6.6.2 Cable types
Power supply: e.g. NYY-J; 3-wire; min. 2.5 mm
2
Analog, signal and transmission cables: e.g. LiYY 10 x 0.34 mm
2
The terminal connection is protected under an additional cover in the top rear section of the
device.
Therefore remove the rear panel of the device to connect the power supply before
commissioning.
The terminal cross-section must be at least 2.5 mm
2
for devices with 24V power supply. With
24V power supply, a current of up to 10A can flow. For this reason pay attention to the
voltage drop on the supply line. The voltage at the device terminals must be within the
specified range (→  54) .