How to troubleshoot incorrect current value in Endress+Hauser Laboratory Equipment?
- MmuellermichealSep 6, 2025
If the current output of your Endress+Hauser Laboratory Equipment shows an incorrect current value, first check the adjustment using the integrated current simulation and connect a mA meter directly to the current output. The load may be too large, or there may be a shunt/short to ground in the current loop.