Liquistation CSF48 System integration
Endress+Hauser 63
8 System integration
8.1 Integrating the sampler in the system
8.1.1 Web server
Versions without fieldbus: An activation code is required for the web server.
Connecting the web server
Connect the communication cable of the computer to the Ethernet port of the BASE2
module.
Base2
Web Browser
A0039619
 64 Web server/Ethernet connection
Establishing the data connection
All versions with the exception of PROFINET:
To ensure that your device has a valid IP address, you must disable the DHCP parameter in the
Ethernet settings. (Menu/Setup/General settings/Extended setup/Ethernet/Settings)
You can assign the IP address manually in the same menu (for point-to-point connections).
All versions including PROFINET:
You will find the IP address and subnet mask of the device under: DIAG/System information/
Ethernet.
1. Start your PC.
2. First, configure a manual IP address in the network connection settings of the operating
system.
Example: Microsoft Windows 10
3. Open Network and Sharing Center.
Apart from your standard network, you should see an additional Ethernet
connection (e.g. as an "unidentified network").
4. Select the link to this Ethernet connection.
5. In the pop-up window select the "Properties" button.