EasyManua.ls Logo

Endress+Hauser Liquistation CSF48 - Page 38

Endress+Hauser Liquistation CSF48
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical connection Liquistation CSF48
38 Endress+Hauser
6.2.1 Wiring the analog inputs and binary inputs/outputs
125
123
124
225
223
224
191
192
291
292
145
146
245
246
1
2
A0042282
 26 Position of the terminals
1 Analog inputs 1 and 2
2 Binary inputs/outputs
6.2.2 Analog inputs
223
225
124
123125
224
Out
In
Gnd
Gnd
In
Out
*
**
*
**
1
2
A0012989
 27 Assignment of analog inputs 1 and 2
* Analog input for passive devices (two-wire transmitter), Out + In terminals (125/123 or 225/223)
** Analog input for active devices (four-wire transmitter), In + Gnd terminals (123/124 or 223/224)