Liquistation CSF48 Commissioning
Endress+Hauser 73
Programming for automatic sampling
In the overview screen, create a simple sampling program under Select sampling program/
New/Basic or under Menu/Setup/Sampling programs/Setup program/New/Basic:
1. Enter the "Program name".
2. The settings from the Basic setup for the bottle configuration and bottle volume are
displayed.
3. Sampling mode=Time paced CTCV is preset.
4. Enter the Sampling interval.
5. Enter the Sampling volume per sample. (For version with vacuum pump, configure
under Menu/Setup/General settings/Sampling.)
6. Select the Bottle change mode after number of samples or time for average samples.
With the option "Bottle change after a time", you can enter the change time and bottle
synchronization (None, 1st bottle change time, 1st time of change + bottle number). The
description for this can be found in the "Bottle synchronization" section.
With the option "Bottle change after a time", you can choose the bottle synchronization
before the start condition (None, 1st bottle change time, 1st time of change + bottle
number). The description for this can be found in the "Bottle synchronization" section.
1. For Multiple bottles enter the number of bottles the sample should be distributed over.
2. Start condition: immediately or after date/time
3. Stop condition: after program end or continuous operation.
4. Pressing SAVE saves the program and ends data entry.