Electrical connection Liquisys M CLM223/253
22 Endress+Hauser
83
84
11
12
13
CYK71
CLK6
15
16
83
84
11
12
13
green
green
white
white
yellow
yellow
brown (nc)
brown (nc)
black
white
red
semiconductor
screen
semiconductor
screen
screen
screen
screen
A0008938-EN
 12 Structure of the special measuring cables
A Semi-conductor layer
sc Shield
For further information on the cables and junction boxes, please refer to the
"Accessories" section.
Field device measuring cable connection
Proceed as follows to connect a conductivity sensor to the field device:
1. Open the housing cover to access the terminal block in the connection compartment.
2. Break through the perforation for a cable gland, mount a cable gland and guide the
cable through this gland.
3. Connect the cable in accordance with the terminal assignment (see connection
compartment sticker).
4. Tighten the cable gland.
Panel-mounted device measuring cable connection