Commissioning Liquisys M CLM223/253
34 Endress+Hauser
7.3 Quick Setup
After power-up, you must make some settings to configure the most important functions
of the transmitter which are required for correct measurement. The following section gives
an example of this.
User entry Range of adjustment
(factory settings in
bold)
Display
1. Press the ENTER key
2. Enter the code 22 to open access to the
menus. Press the ENTER key.
3. Press the MINUS key until you get to the
"Service" function group.
A0008408-EN
4. Press ENTER to be able to make your settings.
5. Select your language in S1, e.g. "ENG" for
English.
Press ENTER to confirm your entry.
ENG = English
GER = German
FRA = French
ITA = Italian
NEL = Dutch
ESP = Spanish
A0008409-EN
6. Press the PLUS and MINUS key simultaneously
to exit the "Service" function group.
7. Press the MINUS key until you get to the
"Setup 1" function group.
A0007824-EN
8. Press ENTER to be able to make your settings
for "Setup 1".
9. In A1, select the desired mode of operation,
e.g. "cond" = conductive.
Press ENTER to confirm your entry.
cond = conductive
ind = inductive
MOhm = resistance
Conc = concentration
A0009002-EN
10. In A2, press ENTER to accept the factory
setting. (Only if A1 = conc, otherwise continue
with Step 12)
%
ppm
mg/l
TDS = total dissolved
solids
None
A0009003-EN
11. In A3, press ENTER to accept the standard
setting.
XX.xx
X.xxx
XXX.x
XXXX
A0009004-EN
12. In A4, press ENTER to accept the standard
setting.
auto, μS/cm, mS/cm,
S/cm, μS/m, mS/m, S/ m